We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BackgroundAnaemia is a deleterious consequence of malaria, and its accurate diagnosis is crucial for effective management. However, laboratory methods for measuring haemoglobin (Hb) concentration, like the Coulter Counter and the Quantitative Buffy Coat® (QBC®), are costly and not widely accessible in resource-limited settings. The point-of-care HemoCue® test is a cheaper alternative and suitable in rural areas. The study aimed to determine the level of agreement between Coulter Counter/QBC® vs. HemoCue®-measured Hb concentrations by Bland-Altman analysis.MethodsAs part of a randomized, placebo-controlled trial of single low-dose primaquine in Ugandan and Congolese children with acute uncomplicated Plasmodium falciparum malaria, Hb concentrations were measured on days 0, 3, 7, and 28 using Coulter Counter (Uganda, n = 1880 paired values), QBC® (DR Congo, n = 1984 paired values) and HemoCue® Hb-301™. The predefined clinically acceptable limits were set at ± 0.5 g/dL.ResultsThe Bland-Altman analysis showed that the HemoCue® minus Coulter Counter mean Hb difference was - 0.15 g/dL with lower and upper limits of agreement of - 3.68 g/dL and 3.39 g/dL, respectively. Corresponding HemoCue® minus QBC® values were - 0.23 g/dL, - 1.66 g/dL and 1.22 g/dL. Linear regression of Hb concentration differences vs. mean Hb concentrations showed negative correlations: r = - 0.43 and r = - 0.34 for HemoCue® vs. Coulter Counter and HemoCue® vs. QBC®, respectively.ConclusionsCompared to Coulter and QBC®, mean HemoCue® measured Hb concentrations were lower and, compared to the Coulter or QBC® methods, had an overall tendency to measure lower Hb concentrations with increasing Hb concentrations. Upper and lower limits of agreement were wider than the predefined clinically acceptable limits of ± 0.5 g/dL. HemoCue® should be used with caution in settings where decisions about blood transfusions are made.

Original publication

DOI

10.1186/s12936-025-05318-5

Type

Journal article

Journal

Malaria journal

Publication Date

03/2025

Volume

24

Addresses

Centre for Tropical Medicine and Global Health, Nuffield Department of Medicine, University of Oxford, Oxford, UK.

Keywords

Humans, Malaria, Falciparum, Anemia, Hemoglobins, Antimalarials, Child, Child, Preschool, Infant, Congo, Democratic Republic of the Congo, Uganda, Female, Male