We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

BACKGROUND: Glucose-6-Phosphate dehydrogenase (G6PD) deficiency is the most common enzymopathy worldwide, no reliable bedside diagnostic tests to quantify G6PD activity exist. This study evaluated two novel quantitative G6PD diagnostics. METHODS: Participants with known G6PD activity were enrolled in Bangladesh. G6PD activity was measured by spectrophotometry, Biosensor (BS; AccessBio/CareStart, USA) and STANDARD G6PD (SG; SDBiosensor, ROK). G6PD activity was measured repeatedly in a subset of samples stored at room temperature and 4°C. RESULTS: 158 participants were enrolled, 152 samples tested by BS, 108 samples by SG and 102 samples were tested by all three methods. In comparison to spectrophotometry BS had sensitivity and specificity of 72% (95%CI: 53-86) and 100% (95%CI: 97-100) at 30% cut off respectively, while SG had a sensitivity of 100% (95%CI: 88-100) and specificity of 97% (95%CI: 91-99) at the same cut off. The sensitivity and specificity at 70% cut off activity were 71% (95%CI: 59-82) and 98% (95%CI, 92-100) respectively for BS and 89% (95%CI: 77-96) and 93% (95%CI: 83-98) respectively for SG. When an optimal cut-off was applied the sensitivity of the BS at 70 cut off rose to 91% [95%CI: 80-96] and specificity to 82% [95%CI: 83-89]; a diagnostic accuracy comparable to that of the SG (p = 0.879). G6PD activity dropped significantly (-0.31U/gHb, 95%CI: -0.61 to -0.01, p = 0.022) within 24 hours in samples stored at room temperature, but did not fall below 90% of baseline activity until day 13 (-0.87U/gHb, 95%CI: (-1.11 to -0.62), p<0.001). CONCLUSION: BS and SG are the first quantitative diagnostics to measure G6PD activity reliably at the bedside and represent suitable alternatives to spectrophotometry in resource poor settings. If samples are stored at 4°C, G6PD activity can be measured reliably for at least 7 days after sample collection.

Original publication

DOI

10.1371/journal.pone.0206331

Type

Journal article

Journal

PLoS One

Publication Date

2018

Volume

13

Keywords

Adult, Enzyme Assays, Enzyme Stability, Female, Glucosephosphate Dehydrogenase, Humans, Male, Point-of-Care Systems